Dissolved Oxygen Analyzer Aqualyzer® 9065
Troubleshooting and Technical Questions
 |
 |
 |
 |
 |
| Nothing is displayed on the
screen when a button is
pressed. |
|
The power supply has not
been connected.
The power supply is in the
safe mode (too hot). |
|
Connect the power supply.
Wait 10 minutes and turn the power
supply back on. |
| The O2 value seems to be
incorrect. |
|
The temperature sensor is
faulty. |
|
Check whether a realistic temperature
value is given.
Check the O2 sensor using calibration
gases and, if necessary, recalibrate it. |
| The message ‘O2 sensor
does not react’ is displayed
on the measurement
screen. |
|
An oxygen sensor has not
been connected.
The oxygen sensor is
faulty. |
|
Connect the oxygen sensor.
Contact Waltron. |
 |
 |
 |
 |
 |
| ‘O2 high’ |
|
O2 value > set O2 ‘high’
alarm |
|
Check the process.
Change the ‘high’ level. |
| ‘O2 high-high’ |
|
O2 value > set O2 ‘high –
high’ alarm |
|
Check the process.
Change the ‘high – high’ level. |
| Sensor x, Temp high alarm |
|
Temp value > set Temp high
value |
|
Check the process
Change the ‘Temp high’ level |
| Sensor x, Temp low alarm |
|
Temp value > set temp low
value |
|
Check the process
Change the ‘Temp low’ level |
| Sensor x, not connected |
|
No sensor is connected
No communication is possible
with the sensor |
|
Connect the sensor |
| Sensor ‘INIT’ error |
|
Not all the calibration data has
been loaded into the control
box |
|
Connect the sensor to the control box and
restart the control box. |
| Sensor x, NTC broken wire |
|
NTC broken |
|
Contact Waltron. |
| Sensor x, short circuit |
|
NTC Short circuit |
|
Contact Waltron. |
| Sensor x, calibration
expired |
|
The sensor has taken more
then 1,000,000 measurements |
|
The sensor must be recalibrated |
| Communication error |
|
No sensor connected
Sensor Broken
Broken wire communication
Cable
Thermal shutdown sensor
(> 65°C) |
|
Check connection
Reset 9065
Cool down the sensor |
symbol appears
in the display |
|
The sensor has taken more
then 1,000,000 measurements. |
|
The sensor must be recalibrated. |