THIS PRODUCT IS DISCONTINUED The Micro 8231 Fluoride Analyzer / Monitor uses a fluoride ion-selective electrode in conjunction with a reference electrode mounted in a temperature controlled flow cell. Pretreatment of the sample with a reagent solution is used to adjust the pH, decomplex the fluoride ions in solution, and remove the effects of changes in ionic strength in the sample. Range of measurement on the fluoride analyzer can be set to any two consecutive decades
of concentration between 0.1 and 1000 ppm. |
![]() |
|
| The current outputs of the fluoride analyzer can be used to view any window within the range to which the instrument is set, e.g. in control of fluoridation a window of 0.1 to 1.5 ppm provides optimum discrimination at the control point, i.e., 1 ppm. Under normal conditions, with suitable standards and calibration frequencies, accuracies better then ±5% of reading or ±0.1 % ppm, whichever is greater, can be achieved. |
| FEATURES | ||
| Fully Automatic Two Point Calibration | Simple to Maintain | |
| High Sensitivity | Highly Reliable Components | |
| Microprocessor Based Ease of Operation | Reduced Reagent Consumption | |
| Complete Technology From One Company | Low Maintenance Liquid Handling | |
| INSTALLATION GUIDELINES | ||
Sample Flow: Suspended Solids: Ambient Temperature: |
Reagent Solutions: Calibration Solutions: |
|
| GENERAL SPECIFICATIONS | ||
Range: Reproducibility: Response Time: Control Temperature Range: Temperature Resolution: Current Output Span:
|
Computer Interface: Alarms: Concentration Alarm Adjustment: Concentration Alarm Differential: Programmable Date Retention: Power Supply: Power Supply Tolerance: Isolation Voltage: Degree of Protection: |
|