9096 Degas Cation Conductivity Analyzer

The NEW Waltron 9096 Degas Cation Conductivity Analyzer delivers accurate measurement of specific, cation, and degassed cation conductivity with unprecedented degassing efficiency of over 94% in under 90 seconds.

During the start up of a thermal power plant, the purity of the generated steam determines whether it can be sent to the steam turbine or if it has to be bypassed to the condenser. Traditionally “Cation Conductivity” instruments are used to analyze steam quality, but in addition to measuring harmful ionic compounds (e.g. Chlorides), they also include CO2, which is not significantly harmful to the steam turbine.

The Waltron 9096 Degas Cation Conductivity Analyzer is based on a patent-pending “Dynamic System”,  meaning it does not use a reboiler or require any supplemental inert gases.

This instrument uses ambient air forced against the flow of the water sample to achieve its high degassing efficiency, resulting in significantly faster indications than any other comparable systems available.

Based on even the most conservative assumptions, the cost savings on each start-up cycle between the Waltron dynamic system and the fastest reboiler system is significant.  Additional benefits are better energy efficiency and reduced emissions of heat and exhaust.

 

 


 Accurate and wide-range analysis


Compact design

Color touchscreen display


Internal datalogger

Low reagent level alarm


No compressed air or bottled gasses required


Automatic start/stop based on sample flow detection


Minimal analyzer maintenance

Wet section can be mounted separately from the transmitter

PERFORMANCE

Range

0-10µS/cm, 0-100µS/cm


Required Maintenance
Monthly or quarterly replenish reagents; Quarterly replace tubing


Reagent Consumption (approximate)
Cation resin (pre-rinsed) and Soda Lime, replace cartridges every two months


Sensor Classification
Cationic exchange and conductivity sensors


Application
10-45°C analyzer (50-113°F)


Ambient Temperature Range
10-45°C analyzer (50-113°F)


Atmospheric Pressure Range
No Limits


Positioning and Installation Details
Surface or separated panel mounting (optional)


Weight
20kg


Degree of Protection
IP55 (NEMA 4): Wet section, IP65 (NEMA 4x): Electronics


Alarms
2 or 4 configurable alarm relays (Result, loss of sample, fault)


Power Supply (on request)
110-220Vac, 50-60 Hz 80 VA


Certification (on request)  
Meets low voltage and low electromagnetic compatibility directives


SAMPLE DELIVERY OPERATING CONDITIONS

Temperature Range
5-55º C (41-131º F)


Inlet Sample
100–500 ml/min, atmospheric, flexible tubing O.D. 1/4"(6mm)


Outlet Sample
Atmospheric, waste tubing O.D. 3/8"(9mm)



SIGNAL OUTPUTS

Analog Outputs
4-20mA (galvanic isolator module available as option) or 0-5V


 

 

Need Assistance?