Highly accurate calculated pH
Alarm for cation resin exhaustion
Easy to replace cation column
Easy to mount stainless steel panel
One transmitter for two conductivity values and pH
Dual channel option (9094C)
The 9095 MultiCon Analyzer delivers accurate measurement of specific conductivity, cation conductivity and calculated pH.
During the start up of a thermal power plant, the purity of the generated steam determines whether it can be sent to the steam turbine or if it has to be bypassed to the condenser. Traditionally “Cation Conductivity” instruments are used to analyze steam quality.
Highly accurate calculated pH
Alarm for cation resin exhaustion
Easy to replace cation column
Easy to mount stainless steel panel
One transmitter for two conductivity values and pH
Dual channel option (9094C)
Range
0-10µS/cm, 0-100µS/cm
Accuracy
< 0.6% of the measuring range
Reagent Consumption (approximate)
Cation Resin (pre-rinsed); replace cartridge every two months
Sensor Classification
Cationic exchange and conductivity sensors
Data Storage
Extract via USB 2.0 (optional)
Power Supply
AC 110 to 240 V +10 / -15%; 48 to 63 Hz or AC/DC 20 to 30 V; 48 t0 63 Hz
Sample Temperature Range
32 – 122° F (0 – 50 °C)
Ambient Temperature Range
32 – 140° F (0 – 60 °C)
Inlet Sample Flowrate
.031 - .044 GPM (7-10L/h), 6mm Swagelok
Inlet Sample Pressure
Maximum of 145 psig (10 bar)
Outlet Sample
Atmospheric
Analog Outputs
5 x 0/4 – 20 mA for measured signals
Alarms
General Alarm, Resin Exhausted Alarm
Optional
RS422/RS485
Construction
Transmitter – high strength ABS, IP67
Wet section – stainless steel or PVC
Dimensions
14.5 x 33.5 x 10" (368 x 851 x 254mm)
Mounting
Wall mount or Panel mount
Finish
Corrosion resistant